- Recall classification
- Class II
Reason stated in the source
The Event Notification Manager (ENM) functionality for certain products does not process notifications for study modifications performed on post verified exams.
Source facts
- Action
- An Urgent Medical Device Correction letter was sent to customers beginning June 14, 2022. The letter informs the customer of the issue and provides safety instructions for continued use of the device. GE Healthcare will correct all affected devices via a software update. In the U.S., this letter will be sent by a traceable means. This letter will be sent to the following titles within the affected accounts: Director/Manager of Radiology, Director/Manager of Cardiology, Risk Manager/Hospital Administrator, Head of Radiology Department, Head of Cardiology Department, PACS Administrator, Director of IT Department, Head of Biomedical Engineering, Head of Imaging Informatics.
- Initiated
- 2022-06-14
- Root cause
- Software design
- Terminated
- 2024-01-16
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CO, NV, OH, PA, WA and the countries of Brazil, Canada, China, France, Germany, Israel, Italy, Japan, Oman, Spain, Switzerland, and United Kingdom.
- Recall number
- Z-1466-2022
- Recall status
- Terminated
- CDRH record ID
- 194640
- FDA record created
- 2022-07-25
- Recall event number
- 90525