Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.
Source facts
Action
The recalling firm alerted affected hospitals by visit on February 24, 2025. Customers were asked to check inventory for affected lot numbers and to return the affected inventory for replacement.
Notifications were issued on February 26, 2025.
Initiated
2025-02-24
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-1467-2025
Recall status
Open, Classified
CDRH record ID
212753
FDA classified
2025-03-26
Recall event number
96387