Z-1468-2025: Progressa Bed Surfaces, intended to be used to treat or prevent pulmonary or other complications associat

Recall record Z-1468-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1468-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Mar 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The air bladders inside the mattress may move out of position when the head of the bed is elevated, causing a dip in the mattress.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 2/25/25 was sent to customers. Actions to be Taken by Customers 1. Please locate all affected units within your facility by reviewing the product label at the head end of the Progressa bed surface with the affected product details provided above. Baxter recommends identifying patients at a higher risk for pressure injuries and placing these patients on alternate beds, if available. 2. A Baxter representative will contact your facility to schedule the correction of your impacted surfaces, or you may contact the Baxter Technical Support Team at 800-445-3720, option 2, between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday. Please work with your Baxter representative to ensure that the affected beds are available for correction (not in use). 3. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply form instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the Biomedical Engineering Department, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and ch
Initiated
2025-02-25
Root cause
Nonconforming Material/Component
Distribution
US Nationwide. Global Distribution.
Recall number
Z-1468-2025
Recall status
Open, Classified
CDRH record ID
212555
FDA classified
2025-03-26
Recall event number
96327

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗