Z-1469-2020: VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Prod

Recall record Z-1469-2020 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1469-2020
Company in FDA record
The Binding Site Group, Ltd.
FDA record creation date
Mar 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Source facts

Action
On XXX, an Urgent Field Safety Notice was sent via email to customers. Customers are instructed to discontinue using this product immediately; dispose of any remaining product at your facility following local regulations; discuss this information with your medical director so that the potential impact of the issue is considered on a local level; and to return the completed and signed E-Back form TSWS18 to Technical.Support@bindingsite.com or to your local firm representative within one week of receiving this notification. Customers are also requested to include a witness signature for the kit disposal as this is required on the E-Back form. Customers with additional questions were encouraged to email Technical.support@bindingsite.com.
Initiated
2018-08-12
Root cause
Under Investigation by firm
Terminated
2022-01-20
Distribution
Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.
Recall number
Z-1469-2020
Recall status
Terminated
CDRH record ID
179633
FDA record created
2020-03-11
Recall event number
84899

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗