Open original FDA record ↗Recall record
- FDA identifier
- Z-1469-2021
- Company in FDA record
- Heartware, Inc.
- FDA record creation date
- Apr 21, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604
Source facts
- Action
- Beginning 26-Feb-2021, an Urgent Medical Device Recall notification was disseminated to consignees who Medtronic records indicate have received affected product, UPS 2-day delivery. The notices provided updated information for the instructions for use and patient manual. Note: this notice covers issues in three separate recalls in FDA's database.
- Initiated
- 2021-02-26
- Root cause
- Device Design
- Distribution
- Worldwide distribution, including: US nationwide, Italy, Great Britain, Germany, Japan
- Recall number
- Z-1469-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186390
- FDA record created
- 2021-04-21
- Recall event number
- 87603