Z-1469-2022: Sight OLO, Automated Hematology Analyzer, Catalog No. OLO-U1

Recall record Z-1469-2022 for SIGHT DIAGNOSTICS LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1469-2022
Company in FDA record
SIGHT DIAGNOSTICS LTD
FDA record creation date
Jul 27, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and incorrectly displayed on OLO devices running software version 2.63, 2.63.1, and 2.63.1.1. Due to the issue, results may incorrectly appear to be within or outside the reference range.

Source facts

Action
On June 14, 2022, the firm notified customers via email of the recall. Customers were informed that a software correction for this issue will be included in any software version ABOVE 2.63.1.1. Pending the software correction through a version upgrade, the firm provided instructions to manually configure the correct default reference ranges on the device, which are also included in Section 2.5.4 of the Sight OLO Operator's Manual. Alternatively, you can contact support@sightdx.com and coordinate a time for Sight's technical support team to remotely update your device software.
Initiated
2022-06-14
Root cause
Software Design Change
Distribution
Worldwide distribution - US Nationwide distribution in the states of Alabama, Arizona, California, Florida, Idaho, Illinois, Kansas, Kentucky, Massachusetts, Michigan, Minnesota, Mississippi, New Jersey, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin and the countries of Burkina Faso, Belize, Canada, Cote D'Ivoire, Cameroon, Curacao, Germany, Spain, Finland, United Kingdom, Greece, Hong Kong, Indonesia, Ireland, Israel, India, Italy, Mexico, Portugal, Qatar, Sweden, Slovenia, Chad, Thailand, South Africa.
Recall number
Z-1469-2022
Recall status
Open, Classified
CDRH record ID
194707
FDA record created
2022-07-27
Recall event number
90533

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗