Z-1469-2023: ARTIS one with a hybrid cable- An angiography system developed for diagnostic imaging and interventional

Recall record Z-1469-2023 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1469-2023
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Apr 27, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Examination room monitor connected with the hybrid cable may have a display issue and no image function is available

Source facts

Action
Siemens issued a Urgent Medical Device Correction letter AX065/22/S on 3/21/23. Letter states reason for recall, health risk and action to take: has been performed. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. Siemens will correct the hardware error via Update Instruction AX061/22/S. This measure will replace the current hybrid cable with a design improvement to enhance stability and quality of the signal transmission. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Initiated
2023-03-21
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of FL, SD, PR and the countries of Angola, Argentina, Australia, Brazil, Bulgaria, China, Georgia, Germany, India, Indonesia, Iraq, Japan, Malaysia, Morocco, Nepal, Nigeria, Poland, Russia, Rwanda, Serbia, Slovenia, South Africa, Spain, Sudan, Syria, Taiwan, Thailand, Uganda, Ukraine, United Kingdom, Uzbekistan, Vietnam.
Recall number
Z-1469-2023
Recall status
Open, Classified
CDRH record ID
199724
FDA classified
2023-04-27
Recall event number
92118

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗