Z-1469-2026: Impella RP with SmartAssist. Product Code: 0046-0035.

Recall record Z-1469-2026 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1469-2026
Company in FDA record
Abiomed, Inc.
FDA classification date
Feb 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.

Source facts

Action
Abiomed notified consignees on about 01/27/2026 via UPS delivery. Additionally, the Customer Letter was posted on Heartrecovery.com: https://www.heartrecovery.com/en-us/resources/downloads/field-safety-notification-impella-rp-differential-pressure-sensor. The notifications informed consignees were instructed clinicians that affected units can continue use and they should: 1. Continue to monitor patient hemodynamics with approved diagnostic devices and continue to verify Impella positioning with imaging before clinical interventions. 2. Refer to the P-level flow rates listed in the Instructions for Use (IFUs), as shown in Figure 1, and rely on these values rather than the AIC display. 3. Trend pump metrics to monitor pump performance and to identify sudden change, noting that abrupt changes in the displayed flow rate need to be assessed both against the flow rate listed in the IFU and across temporal trends. Consignees were instructed that product is NOT being removed from the field and does not need to be returned, patients should be assessed with approved diagnostic devices before clinical interventions, complete and return the provided business reply form (BRF) to Abiomed4437@sedgwick, forward the notification to anyone in the facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products) and if any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
Initiated
2026-01-27
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Canada, Croatia, Czechia, France, Germany, Hong Kong, Italy, Poland, Serbia, Switzerland.
Recall number
Z-1469-2026
Recall status
Open, Classified
CDRH record ID
218097
FDA classified
2026-02-27
Recall event number
98276

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗