Z-1471-2020: GlideScope Video Laryngoscope, Operations & Maintenance Manual, 0900-4940 REV 03

Recall record Z-1471-2020 for Verathon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1471-2020
Company in FDA record
Verathon, Inc.
FDA record creation date
Mar 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A typographical error in the Operations & Maintenance Manual incorrectly identifies the use of low level disinfection for the reusable video laryngoscope, specifically for the Titanium Blades.

Source facts

Action
On 03/31/2019 the firm communicated a "Field Safety Notice Medical Device Recall" letter to customers via FedEx informing them the firm has identified that the disinfection levels outlined in Table 9 (Table 9. System Risk Classification) on page 23 of the Operations & Maintenance Manual (OMM) incorrectly indicates (X in the table cell) that low level disinfection is the minimum disinfection level required for the reusable Titanium video laryngoscope. The correct reference should be an X in the table cell indicating that high-level disinfection is required. The error in Table 9 contradicts the other references and warnings in the OMM that correctly state The reusable Titanium video laryngoscope is considered a semicritical device intended to contact the airway. It must be thoroughly cleaned and undergo high-level disinfection after each use (pages 6, 21, and 34 of the OMM). Customers are instructed to: 1) Determine whether their facility currently has a copy of the In-Service USB (0901-0197) or the GlideScope Video Laryngoscopes Operations and Maintenance Manual 0900-4940 Rev.00 (OMM). If so, dispose of the In-Service USB as well as any printed or saved copies of the OMM. 2) Download and retain the updated GlideScope Video Laryngoscopes Operations and Maintenance Manual 0900-4940 Rev. 01 from the Product Documentation site: https://www.verathon.com/productdocumentation/ 3) Review the disinfection of any reusable Titanium video laryngoscopes completed from January 01, 2019 - February 28, 2019 to ensure that high-level disinfection was completed. High level disinfect any units that were or may have been processed using low-level disinfection. 4) Fill out the attached Recall Response Form and return it by email to CSNotifications@verathon.com. Some facilities may not currently have copies of the In-Service USB (0901-0197) or the OMM subject to this Field Safety Notice (e.g., because they did not retain the In-Service USB or print or save a copy
Initiated
2019-03-31
Root cause
Error in labeling
Terminated
2021-12-14
Distribution
US: RI, IL, FL, SD, NY, MD, CA, OH, LA, AZ, GA, WA, OK, NC, AL, PA, TN, NJ, NV, MI, OR, TX, NM, VA, UT, NE, WI, CO, AR, MN, CT, ND, and KS. OUS: Canada
Recall number
Z-1471-2020
Recall status
Terminated
CDRH record ID
178079
FDA record created
2020-03-11
Recall event number
84329

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗