Open original FDA record ↗Recall record
- FDA identifier
- Z-1472-2020
- Company in FDA record
- Medicrea International
- FDA record creation date
- Mar 11, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.
Source facts
- Action
- On January 28, 2020, the firm notified affected customers of the recall via email. Customers were informed that affected lots may include screws which were incorrectly assembled. Customers were instructed to do the following: - Immediately check your internal inventory and quarantine the recalled product. Document results and coordinate with Medicrea International to return any nonconforming devices identified. - Inform all interested/affected parties. - Inform Medicrea International if any of the recalled devices have been distributed to other organizations. Provide contact details so that Medicrea International can inform the recipients appropriately. Distributors are responsible for retrieving the product. -Complete the customer response form attached to the customer notification even if you no longer have any physical inventory on site.
- Initiated
- 2020-01-28
- Root cause
- Process control
- Terminated
- 2023-03-17
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.
- Recall number
- Z-1472-2020
- Recall status
- Terminated
- CDRH record ID
- 179674
- FDA record created
- 2020-03-11
- Recall event number
- 84928