Z-1472-2026: Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Ad

Recall record Z-1472-2026 for Agiliti Health - Ellis. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1472-2026
Company in FDA record
Agiliti Health - Ellis
FDA classification date
Feb 26, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Source facts

Action
The firm issued an URGENT: Medical Device Correction notice to its consignees on 01/21/2026 via email. The notice explained the issue with the device, potential risk, how to recognize the failure, and requested the following: "Actions to be Taken by Customer 1. Immediately share this Notice with all necessary parties within your organization. 2. Examine the stock in your organization to determine if you have any of the devices in your organization. a. If any devices have been transferred to another organization, please forward a copy of this Advisory Notice and the attached Customer Response Form to that organization. 3. Complete the attached Customer Response Form, even if your organization no longer possesses any devices. 4. Follow the instructions below when using Adapt Line support surfaces to ensure the Microclimate Management function is not inadvertently turned off. Temporary actions Until Agiliti is able to install updated software on devices in your possession to correct the issue, please do the following to ensure that the Microclimate function is restored after using Autofirm. After using Autofirm (if the 10-minute timer elapses) 1. Press the Microclimate Management button to toggle the function on. 2. Verify the indicator light next to the Microclimate Button is illuminated. 3. Verify all previous therapy settings have been restored." For questions regarding this Notice or the Recalled Device, please contact your local Service Representative or the Agiliti customer care team: " Email: AFCA0001@agilitihealth.com " Phone: 800-814-9389, option 3
Initiated
2026-01-21
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-1472-2026
Recall status
Open, Classified
CDRH record ID
218307
FDA classified
2026-02-26
Recall event number
98332

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗