Skip to content

Public FDA record

Z-1473-2020: Iliac Screws, multiple sizes, Article Nos. B36020500 B36020570 B36020580 B36020590 B36027500 B36027570 B

Recall record Z-1473-2020 for Medicrea International. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1473-2020
Company in FDA record
Medicrea International
FDA record creation date
Mar 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.

Source facts

Action
On January 28, 2020, the firm notified affected customers of the recall via email. Customers were informed that affected lots may include screws which were incorrectly assembled. Customers were instructed to do the following: - Immediately check your internal inventory and quarantine the recalled product. Document results and coordinate with Medicrea International to return any nonconforming devices identified. - Inform all interested/affected parties. - Inform Medicrea International if any of the recalled devices have been distributed to other organizations. Provide contact details so that Medicrea International can inform the recipients appropriately. Distributors are responsible for retrieving the product. -Complete the customer response form attached to the customer notification even if you no longer have any physical inventory on site.
Initiated
2020-01-28
Root cause
Process control
Terminated
2023-03-17
Distribution
Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.
Recall number
Z-1473-2020
Recall status
Terminated
CDRH record ID
179873
FDA record created
2020-03-11
Recall event number
84928

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗