Z-1474-04: XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Pho

Recall record Z-1474-04 for Abbott Laboratories HPD/ADD/GPRD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1474-04
Company in FDA record
Abbott Laboratories HPD/ADD/GPRD
FDA classification date
Sep 18, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results.

Source facts

Action
Abbott sent Device Correction letters dated 9/1/04 to all customers receiving the TDx/TDxFLx Benzodiazepines assay, list 9674, informing them of the shift in control values and patient results when used with the XSYSTEMS Dilution Buffer, list 9519-02. The letter gave examples of the effects observed in the benzodiazepines assay performance, and listed the following actions to be taken until Abbott completes their investigation: * During the use of a single lot of buffer, recalibrate the TDx/TDxFLx Benzodiazepines assay every 14 days. * Perform a new calibration run on the TDx/TDxFLx Benzodiazepines assay when a new lot number of XSYSTEMS Dilution Buffer is used. * These actions are in addition to the package insert calibration and quality control requirements. * Follow your laboratory procedures for notification of the health care providers that you serve. Any questions were directed to Abbott''s Customer Service at 1-877-4ABBOTT.
Initiated
2004-09-01
Root cause
Other
Terminated
2004-11-30
Distribution
Nationwide and internationally to Argentina, Australia, Brazil, Chile, Colombia, Costa Rica, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Mexico, Panama, Peru, Puerto Rico, Singapore, South Korea, Taiwan, Thailand, the United Kingdom, Uruguay and Venezuela
Recall number
Z-1474-04
Recall status
Terminated
CDRH record ID
34843
FDA classified
2004-09-18
Recall event number
29957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗