- Recall classification
- Class II
Reason stated in the source
Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
Source facts
- Action
- On 3/2/23 recall notices were mailed to customers who were asked to do the following: 1. Inform individuals within your organization who need to be aware of this action. 2. Identify, remove, and quarantine affected devices. 3. If affected product is found, call customer service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return. 4. Complete and return the Business Reply Form Customers with questions can contact EndoRecall@stryker.com
- Initiated
- 2023-03-02
- Root cause
- Employee error
- Distribution
- Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.
- Recall number
- Z-1474-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199110
- FDA classified
- 2023-04-28
- Recall event number
- 91891