Z-1475-2026: GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046

Recall record Z-1475-2026 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1475-2026
Company in FDA record
GE Healthcare
FDA classification date
Feb 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 01/16/2026 by letter delivered via traceable means. The notice explained the issue and risk, and requested the following: Actions to be taken by Customer/User: "To determine whether you have ZFP Patient Timeline configured at your site, please contact your site administrator. " If you do not have the ZFP Patient Timeline, you can continue to use Centricity Universal Viewer Zero Footprint Client (ZFP) DICOM viewer." "Review all patient studies that were reviewed using the ZFP Patient Timeline to ensure the correct images were reviewed for each patient. GE HealthCare recommends that you discontinue the use of ZFP Patient Timeline." If you do have the ZFP Patient Timeline, pending correction by GE HealthCare, please follow the instructions below: If your site requires the use of Patient Timeline, the following steps must be followed each time a different patient is loaded to ensure the proper series is displayed for this patient: 1. Change the layout to single viewport by selecting the Full view in the Layout Selector (User Manual Section 2.5.1). 2. Manually drag and drop the series from the Navigator into the viewport (User Manual Section 2.4.1). GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2026-01-16
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Belarus, Brazil, Canada, China, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Lebanon, Malaysia, Netherlands, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-1475-2026
Recall status
Open, Classified
CDRH record ID
218475
FDA classified
2026-02-27
Recall event number
98368

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗