Z-1476-2020: QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling an

Recall record Z-1476-2020 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1476-2020
Company in FDA record
Qiagen Sciences LLC
FDA record creation date
Mar 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment

Source facts

Action
QIAGEN issued Urgent Medical Device Correction via email on 1/28/20 advising reason for recall, health risk and action to take: 1) Discontinue use of the Rotor-Gene Q instrument in the following specific conditions: The Rotor-Gene Q instrument is running Software 2.3.4 using the export function to LIMs 2) Upgrade to the latest version of the RGQ software which resolves the issue version 2.3.5. This is expected to be available on 1st February 2020. Please visit the QIAGEN website to upgrade: www.qiagen.com/resources/resourcedetail?id=9d8bda8e-1fd7-4519-a1ff-b60bba526b57&lang=en. Apart from correcting this fault, no other changes to the software have been implemented. Laboratory personnel and clinicians are advised to consider a patients previous test results, other diagnostic tests, anamnesis and current clinical condition when interpreting results from this software. If the results do not match the patients clinical presentation, or incongruences are found with previous and concomitant tests or the results are otherwise unexpected, the patient sample should be retested using an alternate test method or a reference laboratory. 3) Complete the Acknowledgement of Receipt form and return it to QIAGEN. If you have further questions, please contact QIAGEN Technical Services Department on 800-362-7737 (US & Canada) " QIAGEN Subsidiaries: https://www.qiagen.com/about-us/contact/global-contacts/subsidiaries/ " QIAGEN Commercial Partners and Importers: https://www.qiagen.com/about-us/contact/global-contacts/distributors-and-importers/
Initiated
2020-01-28
Root cause
Software change control
Terminated
2024-09-26
Distribution
Worldwide distribution - US Nationwide distribution and countries Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cameroon, Canada, Chile, China, Colombia, Congo, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Laos, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macao, Macedonia, the former Yugoslav Republic of, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mayotte, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestinian Territory Occupied, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab, Republic of Taiwan, Tajikistan, Thailand, Timor-Leste (else known as East Timor), Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic of, Viet Nam, Yemen.
Recall number
Z-1476-2020
Recall status
Terminated
CDRH record ID
179871
FDA record created
2020-03-12
Recall event number
85004

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗