Z-1476-2026: cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material

Recall record Z-1476-2026 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1476-2026
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA classification date
Feb 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.

Source facts

Action
On 1/20/26, the firm notified affected customers via written Urgent Medical Device Correction letters. Customers were provided with workaround instructions and further instructed to consult with the physician or pathologist at their facility to determine specific clinical implications for patients. This issue has been corrected with the launch of a subsequent cobas pro software version. It is mandatory to upgrade all cobas c 503 analytical units running Spline assays to the latest software version. A Roche Diagnostics Service Representative will contact affected customers to schedule an upgrade to software version 03-03.
Initiated
2026-01-20
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-1476-2026
Recall status
Open, Classified
CDRH record ID
218375
FDA classified
2026-02-27
Recall event number
98347

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗