Z-1477-2018: Proteus 235, ProteusPLUS or Proteus ONE produces and delivers a proton beam for the treatment of patients

Recall record Z-1477-2018 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1477-2018
Company in FDA record
Ion Beam Applications S.A.
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When switching from Pencil Beam Scanning treatment mode to Uniform Scanning treatment mode, the tuning setpoints of the scanning magnets are not always taken into account when the proton beam is requested. The mismatch between the setpoint and feedback leads to an error message. Depending on user action, two scenarios are possible: 1. The user resumes the treatment field without analyzing the cause for the error message. This may lead to under-irradiation. 2. The user reboots the Scanning Magnets Power Supply Electronic Unit. This may lead to delay in patient treatment and the patient may need to be re-aligned, which leads to an additional X-ray dose.

Source facts

Action
On January 26, 2018, the firm sent Field Safety Notices to all its consignees by email or delivered personally, alerting them of the issue. The notice recommended that users not resume pauses in Uniform Scanning more than twice without informing on-site firm personnel. The firm also recommended rebooting the Scanning Magnets Power Supply Electronic Unit if the issue occurs. The recalling firm stated it would implement an upgrade of the Scanning Magnet Power Supply Electronic Unit at all affected sites by June 30, 2018. Consignees were instructed to sign and return a copy of a notice read confirmation within 10 working days. Recalling firm Helpdesk Contact: +32 2 507 20 81 (available 24/7)
Initiated
2018-01-26
Root cause
Under Investigation by firm
Terminated
2020-04-27
Distribution
Consignees in four states: VA, IL, NJ, and WA. One consignee located in Germany.
Recall number
Z-1477-2018
Recall status
Terminated
CDRH record ID
162753
FDA record created
2018-10-11
Recall event number
79605

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗