Open original FDA record ↗Recall record
- FDA identifier
- Z-1478-2009
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Jun 26, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Interruption of power-up/boot sequence may cause the paper scale and recorder speed settings of the monitor to revert to factory default may lead to misinterpretation of the fetal trace by the user and to potential incorrect clinical decisions
Source facts
- Action
- Philips Healthcare issued an "Urgent-Medical Device Correction" notification dated May 15, 2009 to Customers in the United States via UPS and notifications outside the US will be managed by Philips representatives in each affected geography. Customers were asked to follow the guidelines in the Action to be Taken by Customer/User section of the Correction Notice until they receive their software upgrade. Questions can be made to Philips response center at 1-800-722-9377.
- Initiated
- 2009-05-12
- Root cause
- Software change control
- Terminated
- 2020-09-23
- Distribution
- Worldwide Distribution
- Recall number
- Z-1478-2009
- Recall status
- Terminated
- CDRH record ID
- 81931
- FDA classified
- 2009-06-26
- Recall event number
- 52026