Inaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation).
Source facts
Action
The firm notified customers via phone alerting them of this affected device, issue, potential hazard, correction, and instructions. The communication also included scheduling an expedited service call for a Field Service Technician to replace the affected Flowmeter Module with a corrected Flowmeter Module. When necessary to avoid delaying or cancelling life-sustaining surgery, users can continue to use the Flowmeter Module while awaiting replacement. Once the issue is recognized by the user, if a replacement Flowmeter Module is available, replacement and reassignment of safety connections of the Flowmeter Module can be accomplished in less than 15 seconds. In the event that a replacement Flowmeter Module is not available, a less common mitigation is the use of a back-up stand-alone centrifugal pump or a stand-alone ultrasonic flowmeter system to provide flow data. Customers should receive a copy of the Urgent Medical Device Recall phone script by e-mail along with a Customer Response Form. The e-mail should be reviewed, and the form completed and returned as indicated. Questions or concerns can be directed Terumo CVS Customer Service at 1-800-521-2818.
Initiated
2018-01-19
Root cause
Employee error
Terminated
2018-07-18
Distribution
Worldwide Distribution -- USA, including the states of IL, KY, and MI; and, the country of Singapore.
Recall number
Z-1478-2018
Recall status
Terminated
CDRH record ID
162847
FDA record created
2018-04-20
Recall event number
79624
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.