Open original FDA record ↗Recall record
- FDA identifier
- Z-1478-2022
- Company in FDA record
- Stryker Instruments
- FDA record creation date
- Jul 28, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.
Source facts
- Action
- On June 24, 2022, the firm notified customers via Urgent Medical Device Recall Letters. Customers were instructed to locate and quarantine any affected products found in their inventory. Upon receiving the customer's completed response form, the firm will provide a shipping label to return any recalled product. Upon receipt of the recalled product, Stryker will contact the customer to arrange for replacement product.
- Initiated
- 2022-06-24
- Root cause
- Process change control
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.
- Recall number
- Z-1478-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194709
- FDA record created
- 2022-07-28
- Recall event number
- 90535