The PedFuse Pedicle Screw Rod Caliper included with the MISquito Pedicle Screw System may be used incorrectly as the Surgical Technique does not adequately convey the intended use of the design.
Source facts
Action
SpineFrontier sent an Urgent Advisory Notice letter to customers via email. The letter identified the affected product, problem and actions to be taken. Customers were asked complete and return the Acknowledgement Form to confirm receipt of the recall notification. For questions call 978-232-3990.
Initiated
2015-03-05
Root cause
Labeling design
Terminated
2015-06-15
Distribution
US Nationwide Distribution in the states of TX, KS, CO, and MA
Recall number
Z-1479-2015
Recall status
Terminated
CDRH record ID
133889
FDA classified
2015-04-21
Recall event number
70488