Open original FDA record ↗Recall record
- FDA identifier
- Z-1479-2022
- Company in FDA record
- Meridian Bioscience Inc
- FDA record creation date
- Jul 28, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.
Source facts
- Action
- On June 22, 2022, the firm notified customers via email. Customers were instructed that if a run is aborted and an error code presented, they SHOULD NOT open the instrument. DO NOT touch the retention ring or the inside of the lid as they may be hot. The firm recommends that customers wait approximately 10 minutes to allow the instrument to cool prior to opening the lid and removing the test devices. The firm has developed a firmware solution to prevent the lid from being opened after an aborted run prior to the completion of the cooling period. The Meridian Bioscience Technical Services team will contact you to set up an appointment for firmware installation once available.
- Initiated
- 2022-06-22
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Chile, Italy, Qatar, and Switzerland.
- Recall number
- Z-1479-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194734
- FDA record created
- 2022-07-28
- Recall event number
- 90549