Open original FDA record ↗Recall record
- FDA identifier
- Z-1479-2023
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- May 23, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube. Siemens Healthineers became aware of one customer site where this issue was identified during clinical operation.
Source facts
- Action
- Siemens Medical Solutions USA, Inc., will provide software updates to the affected systems with software versions VF10 and VF11 with software via Update Instructions (UI) XP007/23/S, XP004/23/S, XP003/23/S, and XP001/23/S
- Initiated
- 2022-11-23
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- US Nationwide - Worldwide Distribution
- Recall number
- Z-1479-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199944
- FDA classified
- 2023-05-23
- Recall event number
- 92229