Open original FDA record ↗Recall record
- FDA identifier
- Z-1479-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Feb 27, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
Source facts
- Action
- On or about December 18, 2025 URGENT: MEDICAL DEVICE CORRECTION - UPDATE letters were sent to customers. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the new Supplemental Guidance. Add a copy of this letter with the Supplemental Guidance to your existing Instruction for Use. You may continue to use the device according to this letter and the instruction for use, which cautions users to ensure that a back-up transducer and probe are sterilized and available prior to beginning a procedure. If you have further distributed this product, identify and forward this notification to them.
- Initiated
- 2025-11-14
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
- Recall number
- Z-1479-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218454
- FDA classified
- 2026-02-27
- Recall event number
- 98359