Skip to content

Public FDA record

Z-1480-2026: EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Recall record Z-1480-2026 for Staar Surgical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1480-2026
Company in FDA record
Staar Surgical AG
FDA classification date
Mar 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Source facts

Action
On 1/21/2026, a Field Safety Notice was mailed to customers who were asked to do the following: 1. If any of the devices identified in the Field Safety Notice Return Response form have not already been implanted, remove them from your inventory and quarantine the product. 2. Complete the Field Safety Notice Return Response form indicating if each device has been implanted or is in your possession. 3. Complete and email the Field Safety Notice Return Response form to customerservice.ag@staar.com. 5. Return affected devices to the firm. Should you have any questions, contact the firm's Representative or Customer Service at customerservice.ag@staar.com or by phone at +41323328888 Monday through Friday, 8:00am to 5:00pm CET.
Initiated
2026-01-21
Root cause
Process control
Distribution
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Recall number
Z-1480-2026
Recall status
Open, Classified
CDRH record ID
218600
FDA classified
2026-03-02
Recall event number
98397

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗