Reason stated in the source
Coloplast is conducting a field correction based on the conclusion that two lots of Restorelle DirectFix Posterior mesh were mislabeled. The incorrect size was printed on the patient label. Specifically the patient label depicts the size as 16x11 cm, however, the correct size is actually 14x15 cm. It is important to note that both the product and box labels were correct.
Source facts
- Action
- Coloplast sent a notification letter dated March 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed their coloplast representative is providing replacement device (s) and new patient labels. For further questions please call 1 (800) 258-3476.
- Initiated
- 2012-03-02
- Root cause
- Error in labeling
- Terminated
- 2012-10-11
- Distribution
- USA ( nationwide) including the states of AL, AZ, AR, CA, CO, FL, GA, ID, IL, KY, MD, MS, MO, NY, PA, TN, VA,WA and WI.
- Recall number
- Z-1481-2012
- Recall status
- Terminated
- CDRH record ID
- 108023
- FDA classified
- 2012-05-02
- Recall event number
- 61411