Z-1481-2022: PRESTIGE ATRAUMATIC GRASPER W/PORT DUAL ACTION, ROTATING 5mm DIAM., 36cm LENGTH. Product Code 8360-10

Recall record Z-1481-2022 for Aesculap Implant Systems LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1481-2022
Company in FDA record
Aesculap Implant Systems LLC
FDA record creation date
Jul 29, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Endoscopic graspers may become separated at the weld from handle to the shaft of the device.

Source facts

Action
A notification letter was issued April 30, 2020 sent USPS Certified mail . The notification identifies the products as Catalog Item # 8360-10 PRESTIGE ATRA GRASPER DBL-ACT 5MM and 8361-10 PRESTIGE RETR GRASPER DBL-ACT 5MM. Customers were to evaluate inventory with a "tug test" and submit the notification acknowledgment form. Should a device fail the tug test, customers were to file a complaint using Aesculap's Online Customer Complaint Submission Form and return a the product. The products have since been discontinued.
Initiated
2020-04-01
Root cause
Process change control
Distribution
United States Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Washington DC and Puerto Rico.
Recall number
Z-1481-2022
Recall status
Completed
CDRH record ID
194589
FDA record created
2022-07-29
Recall event number
90493

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗