Open original FDA record ↗Recall record
- FDA identifier
- Z-1481-2022
- Company in FDA record
- Aesculap Implant Systems LLC
- FDA record creation date
- Jul 29, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Endoscopic graspers may become separated at the weld from handle to the shaft of the device.
Source facts
- Action
- A notification letter was issued April 30, 2020 sent USPS Certified mail . The notification identifies the products as Catalog Item # 8360-10 PRESTIGE ATRA GRASPER DBL-ACT 5MM and 8361-10 PRESTIGE RETR GRASPER DBL-ACT 5MM. Customers were to evaluate inventory with a "tug test" and submit the notification acknowledgment form. Should a device fail the tug test, customers were to file a complaint using Aesculap's Online Customer Complaint Submission Form and return a the product. The products have since been discontinued.
- Initiated
- 2020-04-01
- Root cause
- Process change control
- Distribution
- United States Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Washington DC and Puerto Rico.
- Recall number
- Z-1481-2022
- Recall status
- Completed
- CDRH record ID
- 194589
- FDA record created
- 2022-07-29
- Recall event number
- 90493