Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.
Source facts
Action
Siemens Medical Solutions USA, Inc., will provide software updates to the
affected systems with software versions VF10 and VF11 with software via Update
Instructions (UI) XP007/23/S, XP004/23/S, XP003/23/S, and XP001/23/S
Initiated
2022-11-23
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide - Worldwide Distribution
Recall number
Z-1481-2023
Recall status
Open, Classified
CDRH record ID
199948
FDA classified
2023-05-23
Recall event number
92229