Skip to content

Public FDA record

Z-1483-2022: Medline Sprinter OTW, Model/Catalog Number SPR2512W; The Sprinter Over-the-Wire balloon dilatation cathet

Recall record Z-1483-2022 for Medtronic Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1483-2022
Company in FDA record
Medtronic Inc
FDA record creation date
Aug 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Luer has balloon diameter printed as "2.25mmx12mm" instead of the correct diameter, "2.5mmx12mm" as reflected on the product outer packaging.

Source facts

Action
The firm issued a recall notice on 06/24/2022 by US Mail. Instruction to Customers: 1. Identify and quarantine all unused affected products. 2. Return all unused affected products in your inventory to Medtronic for replacement product or destroy all unused affected products in your inventory at your location. To initiate a product return contact Medtronic Customer Service at 800-854-3570. You local Medtronic Representative can assist you as necessary in initiating the return of this product. 3. Complete the enclosed Customer Confirmation Form and email it to rs.cfqfca@medtronic.com. 4. Consignees are asked to forward the notification to those who need to be aware within the organization and to any location which the devices have been distributed.
Initiated
2022-06-24
Root cause
Labeling Change Control
Distribution
US Nationwide distribution.
Recall number
Z-1483-2022
Recall status
Open, Classified
CDRH record ID
194755
FDA record created
2022-08-01
Recall event number
90552

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗