Open original FDA record ↗Recall record
- FDA identifier
- Z-1484-2024
- Company in FDA record
- Fresenius Kabi USA, LLC
- FDA classification date
- Apr 11, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.
Source facts
- Action
- A letter titled URGENT - Medical Device Field Correction, dated 3/7/24, was delivered by hand to consignees or by mail notifying them of this recall event. This notice details eleven software anomalies that could pose a potential risk of serious patient harm or death and the mitigations consignees should perform until software is updated. Consignees are to update their software to LVP software version 5.9.1 by reaching out to their Fresenius Kabi representative at 1-978-775-8100 to schedule installation. Consignees are to complete and return the Customer Reply Form via fax or email. Questions about the recall notice can be directed to Fresenius Kabi by email at Ivenix_support@fresenius-kabi.com or by phone at 1-978-775-8100 (Monday through Friday from 8:30 am to 5:00 pm).
- Initiated
- 2024-03-07
- Root cause
- Software Design Change
- Distribution
- US Distribution: CA, CO, ID, MI, NE, NJ, NV, OK, TX, UT, VA, WA, & WI.
- Recall number
- Z-1484-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206605
- FDA classified
- 2024-04-11
- Recall event number
- 94228