Open original FDA record ↗Recall record
- FDA identifier
- Z-1486-2020
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Mar 13, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Source facts
- Action
- On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.
- Initiated
- 2019-12-30
- Root cause
- Component design/selection
- Terminated
- 2022-11-17
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Recall number
- Z-1486-2020
- Recall status
- Terminated
- CDRH record ID
- 179271
- FDA record created
- 2020-03-13
- Recall event number
- 84756