On 02/26/2021, it was found that there were incorrectly packaged Dilator sets in Finished Goods. On 03/1/2021, it was confirmed that 14 kits containing the impacted products had been distributed to the field.
Source facts
Action
On 3/11/2021 LivaNova issued an Urgent Medical Device Correction notice via letter notifying customers the firm identified an error with the assembly of a limited number of ProtekSolo Venous Dilator Sets. The ProtekSolo Venous Dilator Set 14-22 Fr should contain 3 dilators: 14Fr, 18Fr and 22Fr. A limited number of boxes labeled as ProtekSolo Venous Dilator Set 14-22 Fr incorrectly contain 4 dilators: 14Fr, 18Fr, 22Fr and 26Fr.
Initiated
2021-03-19
Root cause
Under Investigation by firm
Terminated
2021-11-22
Distribution
US Nationwide distribution in the states of AL, AZ, IL, TN.
Recall number
Z-1486-2021
Recall status
Terminated
CDRH record ID
186425
FDA record created
2021-04-23
Recall event number
87632