Z-1486-2025: GE Healthcare Centricity Cardiology CA1000 (CA1000), Model Numbers: 1) 2033901-001; 2) 2038437-001; 3

Recall record Z-1486-2025 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1486-2025
Company in FDA record
GE Healthcare
FDA classification date
Mar 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification. Please ensure you have implemented the actions below: 1: Ensure your hospital network is secure and 2: Outside of the hospital network implement additional safeguards such as a VPN. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.85480@gehealthcare.com." "If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative." Per status report 05/23/2025, the firm send a follow-up notice dated 05/13/2025 providing an installation manual addendum that provides instructions for changing passwords.
Initiated
2025-02-19
Root cause
Software design
Distribution
Worldwide distribution.
Recall number
Z-1486-2025
Recall status
Open, Classified
CDRH record ID
212700
FDA classified
2025-03-28
Recall event number
96375

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗