Seven 18mm x 18mm x 18 mm implants had been placed in kits, packed and labeled as 18mm x 14mm x 14mm.
Source facts
Action
The firm notified their distributors by email on 02/16-17/2017. The distributors were asked to return the unused lot #101866. The firm also contacted the implanting surgeons either by visit or telephone about the potential for an incorrect size of the implant and asked if there were any adverse events.
Initiated
2017-02-16
Root cause
Labeling mix-ups
Terminated
2017-04-27
Distribution
Al, FL, IN, Il, NE, TX, NM, CO, AZ
Recall number
Z-1487-2017
Recall status
Terminated
CDRH record ID
153844
FDA classified
2017-03-20
Recall event number
76687