Z-1487-2021: Spectrum IQ - Infusion Pump - Product Usage: intended to be used for the controlled administration of flu

Recall record Z-1487-2021 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1487-2021
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Apr 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potentially defective component in the AC power adapter which may fail over time.

Source facts

Action
On 3/19/2020 the firm sent a letter to its customers with the following instructions: 1. Operators may continue to use the Spectrum IQ Infusion Pumps until the correction is completed. If using an impacted pump to infuse critical medication such as vaso-active medications, ensure there is a back-up pump available in case the power adapter fails. 2. A local Baxter service representative will contact your facility to provide a list of affected serial numbers at your facility and to coordinate power adapter replacement. Your facility will be receiving this power adapter from Baxter at no charge. 3. Upon receipt of the new power adapters, refer to the Spectrum IQ Installation and Maintenance Manual (41019v900 Rev F, Section 11.2) for instructions on how to remove and replace the power adapter on the impacted pumps. The manual can be found on Baxters Technical Service Portal at https://service.baxter.com. 4. After the installation of the new power adapters, please return the original power adapters to Baxter using the packaging from the replacement power adapter and the provided return label. 5. If you received this notification directly from Baxter, complete the enclosed Baxter Customer Reply Form and return it to Baxter by e-mailing it to fca@baxter.com. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 6. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 7. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 8. If you are a de
Initiated
2021-03-19
Root cause
Device Design
Terminated
2022-07-15
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CT, FL, IA, IN, LA, MA, MD, ME, MN, MO, MS, MT, NC, NJ, NY, OH, PA, RI, SC, TN, TX and the country of Puerto Rico.
Recall number
Z-1487-2021
Recall status
Terminated
CDRH record ID
186307
FDA record created
2021-04-23
Recall event number
87568

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗