Open original FDA record ↗Recall record
- FDA identifier
- Z-1488-2021
- Company in FDA record
- Luminex Corporation
- FDA record creation date
- Apr 23, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for a false-negative result for Vibrio parahaemolyticus, Salmonella, and/or Shigella spp when used with the VERIGENE Enteric Pathogens Nucleic Acid Test Kit.
Source facts
- Action
- A customer advisory notice dated 03/18/2021. The Customer Advisory Notice recommends customers review their QC data to ensure passing results for all VERIGENE EP test targets, and to contact Luminex if any false negatives are observed for any targets within the past twelve months, in order to arrange for a replacement of the Amplification Heater. Additionally, if a Processor SP unit is indicated to have received service 15 on the TSB Log, the Amplification Heater would have already been replaced under routine maintenance with an in specification Amplification Heater and no further action is required for that instrument.
- Initiated
- 2021-03-18
- Root cause
- Process control
- Distribution
- Worldwide distribution - United States (nationwide) and the countries of Austria, France, Greece, Japan, Netherlands, Turkey.
- Recall number
- Z-1488-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186372
- FDA record created
- 2021-04-23
- Recall event number
- 87591