Skip to content

Public FDA record

Z-1488-2021: VERIGENE Processor SP - Product Usage: intended for processing and identifying nucleic acids utilizing go

Recall record Z-1488-2021 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1488-2021
Company in FDA record
Luminex Corporation
FDA record creation date
Apr 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for a false-negative result for Vibrio parahaemolyticus, Salmonella, and/or Shigella spp when used with the VERIGENE Enteric Pathogens Nucleic Acid Test Kit.

Source facts

Action
A customer advisory notice dated 03/18/2021. The Customer Advisory Notice recommends customers review their QC data to ensure passing results for all VERIGENE EP test targets, and to contact Luminex if any false negatives are observed for any targets within the past twelve months, in order to arrange for a replacement of the Amplification Heater. Additionally, if a Processor SP unit is indicated to have received service 15 on the TSB Log, the Amplification Heater would have already been replaced under routine maintenance with an in specification Amplification Heater and no further action is required for that instrument.
Initiated
2021-03-18
Root cause
Process control
Distribution
Worldwide distribution - United States (nationwide) and the countries of Austria, France, Greece, Japan, Netherlands, Turkey.
Recall number
Z-1488-2021
Recall status
Open, Classified
CDRH record ID
186372
FDA record created
2021-04-23
Recall event number
87591

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗