Open original FDA record ↗Recall record
- FDA identifier
- Z-1489-2014
- Company in FDA record
- Ev3, Inc.
- FDA classification date
- Apr 23, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Covidien is recalling 2 production lots of EverCross 0.035 PTA Balloon Catheter because of a product labeling error. There is a discrepancy in the PTA Balloon length and the delivery catheter length printed on the product labeling.
Source facts
- Action
- Consignees were sent a Covidien "Urgent Product Recall" letter dated April 09, 2014. The letter described the problem and the product involved in the recall. The letter described the required actions which included to immediately quarantine and discontinue use of the affected devices and to return affected product. Requested consignees to complete the Verification Form and return it to their Covidien Sales Representative. For questions they can contact Covidien Customer Service at 1-800-716-6000.
- Initiated
- 2014-04-09
- Root cause
- Labeling Change Control
- Terminated
- 2014-10-06
- Distribution
- Worldwide Distribution-USA including the states of WV, TX, OK, IN, NY, TX, and NC, and the countries of AUSTRALIA, ISRAEL, TURKEY, UNITED KINGDOM, ITALY, FRANCE, and CHINA.
- Recall number
- Z-1489-2014
- Recall status
- Terminated
- CDRH record ID
- 126834
- FDA classified
- 2014-04-23
- Recall event number
- 68034