Z-1489-2021: VITROS Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ too

Recall record Z-1489-2021 for Ortho-Clinical Diagnostics, INc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1489-2021
Company in FDA record
Ortho-Clinical Diagnostics, INc.
FDA record creation date
Apr 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.

Source facts

Action
On 3/25/2021, Ortho Clinical Diagnostics issued an Urgent Medical Device Correction notice via letter notifying customers two lots of the VITROS Black Reference Slides have the incorrect part number on the box.
Initiated
2021-03-26
Root cause
Under Investigation by firm
Terminated
2023-05-24
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV, WY and the countries of Australia, Bermuda, Brazil, Canada, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.
Recall number
Z-1489-2021
Recall status
Terminated
CDRH record ID
186458
FDA record created
2021-04-23
Recall event number
87649

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗