Recall record Z-1490-2019 for Change Healthcare Israel Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1490-2019
Company in FDA record
Change Healthcare Israel Ltd.
FDA record creation date
May 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

users are not notified of procedure medication discrepancies between the Vitals and Meds, the Procedure Notes and Patient Common Data screens in Horizon/McKesson Cardiology Hemo" (Hemo). The discrepancy may also affect administered medication data in reports generated from Hemo or the Horizon/McKesson Cardiology Physician s Report as well as customers, who have implemented an outbound interface of procedure results.

Source facts

Action
Change Healthcare notified consignees on about 03/25/2019 via "Urgent Field Safety Notice" letter sent through email. The notification informed consignees that the issue can occur under the following 3 circumstances: 1. When a user intentionally updates the Physician Report, overriding procedure medication derived from the Hemo report, or 2. When a user intentionally updates the medication information charted for Pre- or Intra- procedure stage, from the Intra- or Post-procedure stage (respectively), or 3. Due to an unexpected system error. Instructions included that prior to closing a Hemo case, ensure the list of administered medication information is correct on the Vitals and Meds, Procedure Notes, and Patient Common Data screens, and when opening, finalizing or closing the Physician Report, review the procedure medications to ensure that all medication information is correct. Customers were requested to immediately contact Change Healthcare to acknowledge they have read and understand the Urgent Field Safety Notice and to arrange for installation of the software update/upgrade. They were also requested to ensure distribution of the Urgent Field Safety Notice to all personnel within your organization who should be aware of the recall or to other affiliated parties that may be affected by the Urgent Field Safety Notice. Customer Support contacts are as follows: 1. US/Canada: 1-877-654-4366; 2. UK / Ireland: +44 (0) 208 9527399; 3. Australia: 1800 625 435; 4. Israel: +972 52 433 3366
Initiated
2019-03-25
Root cause
Software Design Change
Terminated
2025-03-04
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, WY. International distribution to Australia, Canada, Germany, Ireland, Israel, United Kingdom.
Recall number
Z-1490-2019
Recall status
Terminated
CDRH record ID
172011
FDA record created
2019-05-24
Recall event number
82600

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗