Skip to content

Public FDA record

Z-1490-2026: Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038

Recall record Z-1490-2026 for ESAOTE S.P.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1490-2026
Company in FDA record
ESAOTE S.P.A.
FDA classification date
Mar 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.

Source facts

Action
Esaote issued a Field Safety Notice to its consignees on 3/14/2025 via FedEx. The notice explained the problem with the device, potential risk of use and requested the following: "Actions to be performed by the user " If the operator detects the issue during the inspection, they must NOT use the probe and should contact Esaote's technical support. " Please distribute this Safety Notice to all personnel involved within your organization and to all users of the device. " If the device is no longer in your possession, please provide Esaote with the new owner's details or inform us about the current status of the device. As a precautionary measure, and to complete the root cause analysis while improving our processes and customer service, Esaote will proceed with a voluntary precautionary recall. Our Technical Support Service will soon contact customers who own the affected units to arrange for the replacement of the probe."
Initiated
2025-03-14
Root cause
Process control
Distribution
US: Nationwide distribution in the states of NC and TX.
Recall number
Z-1490-2026
Recall status
Completed
CDRH record ID
218322
FDA classified
2026-03-03
Recall event number
98339

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗