Open original FDA record ↗Recall record
- FDA identifier
- Z-1490-2026
- Company in FDA record
- ESAOTE S.P.A.
- FDA classification date
- Mar 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.
Source facts
- Action
- Esaote issued a Field Safety Notice to its consignees on 3/14/2025 via FedEx. The notice explained the problem with the device, potential risk of use and requested the following: "Actions to be performed by the user " If the operator detects the issue during the inspection, they must NOT use the probe and should contact Esaote's technical support. " Please distribute this Safety Notice to all personnel involved within your organization and to all users of the device. " If the device is no longer in your possession, please provide Esaote with the new owner's details or inform us about the current status of the device. As a precautionary measure, and to complete the root cause analysis while improving our processes and customer service, Esaote will proceed with a voluntary precautionary recall. Our Technical Support Service will soon contact customers who own the affected units to arrange for the replacement of the probe."
- Initiated
- 2025-03-14
- Root cause
- Process control
- Distribution
- US: Nationwide distribution in the states of NC and TX.
- Recall number
- Z-1490-2026
- Recall status
- Completed
- CDRH record ID
- 218322
- FDA classified
- 2026-03-03
- Recall event number
- 98339