Z-1491-2026: Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF

Recall record Z-1491-2026 for Diagnostica Stago, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1491-2026
Company in FDA record
Diagnostica Stago, Inc.
FDA classification date
Mar 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

Source facts

Action
On February 4, 2026 URGENT: MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers. Actions required: If you currently have one or more kits of STA LIATEST FREE PROTEIN S 6 : - For REF 00516: Lots 271971, 272859, and 273018 with a short expiration date we ask you to discontinue their use and dispose of them in accordance with your local regulations. Stago will provide a credit for any destroyed kits. - For REF 00516: lot 273426: o If the initial barcode sheet provided with the kit has already been loaded into the instrument: - Stop the use the initial barcode pre-calibration settings. - Replace the initial barcode settings with the new pre-calibration barcode settings attached, by following to the procedure described in Appendix 1 of this letter. o If the initial barcode sheet has not yet been loaded into the instrument: - Do not use it and proceed to load the lot using the new pre-calibration barcodes provided in the attachment.
Initiated
2026-02-04
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MI, MN, MO, MT, NC, ND, NJ, NY, OH, PA, TN, TX, UT, VA, WA and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BIELORUSSIE, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, GUATEMALA, INDIA, IRAQ, IRELAND, ITALY, JORDAN, KENYA, KOREA REPUBLIC OF, LITUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PALESTINIAN TERRITORY, PARAGUAY, PORTUGAL, ROMANIA, RUSSIE, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIE, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY;
Recall number
Z-1491-2026
Recall status
Open, Classified
CDRH record ID
218594
FDA classified
2026-03-03
Recall event number
98394

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗