Skip to content

Public FDA record

Z-1492-2014: Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116

Recall record Z-1492-2014 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1492-2014
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
Apr 24, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).

Source facts

Action
Ion Beam Applications ( IBA) sent a recall notification letter dated on/about December 5, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Short term action: A user notice (see annex 1) has been sent to impacted sites where the affected legacy PTS software version was delivered. This user notice explains the workaround required to be put in place in order to fully mitigate the risk and the 1BA action plan to permanently fix this issue. Corrective action: A patch on the PTS Legacy SW was requested and in implementation. In the same time, the solution was put in main code of PTS legacy software version 7 to be integrated in the next release. The customer representative signing underneath confirms that the user notice above is communicated to the appropriate employees. We apologize in advance for any inconvenience this may cause and we would like to thank you for your cooperation. The IBA Site Staff or the Operations manager is available to provide you with additional information and/or guidelines if necessary. Further questions please call : 215-972-7777
Initiated
2013-12-04
Root cause
Software design
Terminated
2014-06-12
Distribution
US Distribution including the states of OK, IL, NJ, and WA.
Recall number
Z-1492-2014
Recall status
Terminated
CDRH record ID
126051
FDA classified
2014-04-24
Recall event number
67678

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗