Skip to content

Public FDA record

Z-1492-2021: EXALT Controller - Product Usage: intended for use with a Boston Scientific endoscope for endoscopic diag

Recall record Z-1492-2021 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1492-2021
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Apr 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

May have been programmed with a less than optimal brightness setting resulting in a darker than desired image

Source facts

Action
Boston Scientific issued Medical Device Removal - Immediate Action Required letter on 3/16/21. Letter states reason for recall, health risk and action to take: While no adverse health consequences are anticipated and an impacted Controller can continue to be used, replacement of the Controller is advised for optimal image quality. A Boston Scientific representative will be in contact with you directly to schedule installation of a replacement Controller at your facility. If you notice a darker than desired image, manually adjust the brightness setting on the front of the EXALT" Controller to the High setting (all five bars on the Brightness Indicator illuminated) A Boston Scientific representative will be in contact with you directly to schedule the installation of a replacement Controller at your facility. Your current Controller will be removed at that time. Please read carefully through the enclosed instructions. We ask that you please confirm receipt of this notification by completing the enclosed Acknowledgement Form and sending it to BSCFieldActionCenter@bsci.com or fax to 763-415-7708.
Initiated
2021-03-16
Root cause
Process control
Terminated
2022-05-02
Distribution
US Nationwide distribution in the states of CA, FL, NJ.
Recall number
Z-1492-2021
Recall status
Terminated
CDRH record ID
186375
FDA record created
2021-04-26
Recall event number
87593

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗