During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Source facts
Action
Terumo Cardiovascular issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 03/03/2025 via email. The notice explained the reason for the removal, potential hazard, and requested the following:
1. Review this Medical Device Correction and assure that all users have received notice of this issue.
2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form.
3. Terumo CVS will contact users to discuss removal and correction activities.
Note: Terumo CVS recommends that users continue using CDI OneView System H/S Probes while waiting for this removal
Initiated
2025-03-03
Root cause
Software design
Distribution
US: IL, TX, NY, MI, FL, NV, and Australia, New Zealand
Recall number
Z-1492-2025
Recall status
Open, Classified
CDRH record ID
212566
FDA classified
2025-03-31
Recall event number
96331