Skip to content

Public FDA record

Z-1492-2026: Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012

Recall record Z-1492-2026 for Vascutek, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1492-2026
Company in FDA record
Vascutek, Ltd.
FDA classification date
Mar 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.

Source facts

Action
On 12/18/2025, Field Safety Notices were emailed to customers who were asked to do the following: Awareness of this Field Safety Notice is advised and shall be distributed to operating surgeons and users. The shelf life ranges (expiry dates) have now been corrected for all newly manufactured products. Please share this information with anyone in your organization who needs to be aware or is a user of the affected devices. Complete and return the Field Safety Notice Return Confirmation form to: taukfsn2025-003shelflife@terumoaortic.com Questions can be sent to the this email address as well.
Initiated
2025-12-18
Root cause
Error in labeling
Distribution
US Nationwide distribution including in the states of FL, PA, TX, NC, NY, CA, NJ, OH, MI, DC, CO, VA, GA, IL, OR, ND, VT, MS, MA, IN, WA, AZ, MN, MD, IA, OK, MO, AL, WI, LA, AR, TN, CT, ME, NE, SC, KS, NV, KY, RI, WV, UT, MT, ID, AK, NH, DE, NM, SD, HI, WY.
Recall number
Z-1492-2026
Recall status
Open, Classified
CDRH record ID
218319
FDA classified
2026-03-04
Recall event number
98338

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗