Open original FDA record ↗Recall record
- FDA identifier
- Z-1493-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- May 7, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Siemens became aware of a potential safety risk related to the behavior of the 160 MLC when using the "Motion Stop" button followed by "Reset" several times during the treatment of one beam or one segment (issue 1). Additionally, there may be a potential for collision if the operator does not ensure clearance of the extended flat panel positioner for all gantry rotation angles (issue 2).
Source facts
- Action
- Siemens sent a Safety Advisory Notice with Update to all affected customers. The notice informed the customers of the risk of pressing the Motion Stop button repeatedly during treatment. Work Steps were provided for customers to follow.
- Initiated
- 2012-04-03
- Root cause
- Labeling design
- Terminated
- 2012-07-25
- Distribution
- Worldwide Distribution - USA including the states of CA, DE, FL, IL, KY, LA, MA, MI, NE, NJ, NY, OH, PA, SD, TN, UT, WI, and WV
- Recall number
- Z-1493-2012
- Recall status
- Terminated
- CDRH record ID
- 108483
- FDA classified
- 2012-05-07
- Recall event number
- 61580