Skip to content

Public FDA record

Z-1493-2019: TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China)

Recall record Z-1493-2019 for Abbott Laboratories Inc. (St Jude Medical). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1493-2019
Company in FDA record
Abbott Laboratories Inc. (St Jude Medical)
FDA record creation date
Apr 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.

Source facts

Action
The recall notice, and clinician acknowledgment form were hand delivered by Abbott Sales Representatives to all impacted accounts starting April 5, 2019. Completed Clinician Acknowledgement Forms, signify they received the correction notification for the impacted TactiSys Quartz Equipment with software version 1.7.
Initiated
2019-04-05
Root cause
Software design
Terminated
2021-04-27
Distribution
US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA
Recall number
Z-1493-2019
Recall status
Terminated
CDRH record ID
172566
FDA record created
2020-04-09
Recall event number
82727

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗