The incorrect dental pins were packaged and subsequently distributed.
Source facts
Action
On 3/18/2021, Coltene issued an Urgent Medical Device Recall notice via letter to customers notifying them the incorrect dental pins were packaged and subsequently distributed. The TMS LINK PLUS Self-Threading L741 Pins should be stainless steel, single sheer link pins that are gold-plated. The pins that were packed and distributed are titanium, double sheer, and not gold plated. The incorrectly packaged pin is too small for the drill, resulting in a hole that may be drilled too large for the pins. The pins may not engage the hole to effectively sheer.
Initiated
2021-03-18
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of NJ, IN, NC, IL. GA, CA, NY, PA and the countries of Canada, Venezuela.
Recall number
Z-1494-2021
Recall status
Open, Classified
CDRH record ID
186441
FDA record created
2021-04-26
Recall event number
87645
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.