Skip to content

Public FDA record

Z-1495-2021: Artis zee ceiling-interventional Fluoroscopic X-Ray System Model Number: 10094137

Recall record Z-1495-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1495-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
May 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some Artis zee ceiling systems show an increased abrasion of the cabling at the cable outlet of the inner C-arm, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.

Source facts

Action
Siemens issued Urgent Medical Device Correction letter dated March 26, 2021. Letter states reason for recall, health risk and action to take: Siemens will inspect the potentially affected area and will install an additional cover to improve the cable guidance and prevent further cable harness damage via Update Instruction AX008/20/S. If damage caused by abrasion is too severe, the cable harness will be replaced via Update Instruction AX011/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Initiated
2021-03-26
Root cause
Device Design
Distribution
Nationwide
Recall number
Z-1495-2021
Recall status
Open, Classified
CDRH record ID
186738
FDA record created
2021-05-13
Recall event number
87672

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗